MTC2
surge tester at 6, 12, 15, 25, 30, 40, or 50 kV
Discontinued – This device is no longer available for purchase. However, we will continue to provide service for your existing devices.
A new generation of this device is now available: MTC2R7
The MTC2's wide range of features allows for a detailed view of the winding.
With our state-of-the-art surge tester Test you can reliably obtain stators, armature, coils, and all types of winding materials.
Perfectly complemented by resistance, partial discharge, insulation resistance, and high voltage test.
The MTC2 – ideal for laboratories, workshops, and automation.
Over 40 years of pioneering spirit, diverse expertise, continuous optimization, and the R&D efforts of our engineers and technicians form the foundation of our surge tester
Designed & Made in Germany.
- Class-leading features
- surge test up to 50 kV
- Surge current 2,000 A
- surge power 125 joules
- Rise time ≥ 60 ns
- Partial discharge testing in accordance with standards
- Suitable for motors / generators up to 500 MW
- Unique variety of analytical methods
- Automatic test method and test terminal switching
- Resistance test four-wire technology , with temperature compensation
- Insulation resistance and PI testing
- Inductance test
- Anchor Test Assistant, Anchor Adapter, and armature booster
- Up to 5 test terminals | 4 x windings, 1 x stator-/motor housing
- Quick manual " Test " without a test plan
- Automatic testing with automated pass/fail comparison
- Integrated industrial PC with Windows10®
- Test report printing plus PDF generation
- Print to any Windows printer
- memory with no limit on test plans and test results
- Full network connectivity
- Intuitive operation via touchscreen, mouse and keyboard
- Remote maintenance and remote calibration
- Ideal for OEMs – remotely controllable for automation

Intuition – Function – Perfection.
The MTC2 is our surge tester top-of-the-line model. No other testing device offers such a wide range of capabilities. With the MTC2, you can test coils, stators, armature , and all types of wound products using state-of-the-art technology—reliably and precisely.
Our patented surge stress analysis method is the basis of our technologically advanced solution. The 6th generation MTC2 surge voltage testers are available in a finely graded variety of devices from 6 kV to 50 kV .
MTC2 – surge tester the luxury class
» 5 Test Leads
5 test leads allow you to connect to all relevant winding terminals at once. The 3 phase terminals and the neutral point (if present and accessible) are connected using 4 test leads. The fourth test connection at the neutral point provides a significant increase in test accuracy. The fifth test connection is connected to the stator-/motor housing. It is required for the insulation test between the winding and the laminated core .
» automatic test method switch-over
After connecting the test object, the MTC2 automatically tests on all 5 terminals. Thanks to the integrated test method switching, you no longer have to reclamp connections by hand. This saves time and avoids errors.
The surge test with standard-compliant flank steepness ensures that no defect goes undetected. Modern automated evaluations make the subsequent defect analysis easier for you.
" Partial Discharge Testing at surge test " (optional)
The standard-compliant PD test can also be accessed at built-in . This provides information about critical insulation weaknesses and suitability for operation with frequency converters. Here, SCHLEICH employs two strategies to ensure accurate measurement results both on the stator and on the fully assembled Motor :
- Motors are measured using a coupler in the test lead .
- Stators are measured with the help of an antenna.
» Precise resistance measurement in four-wire technology (optional)
To obtain highly accurate measurement results for your resistance comparison between the three phases, the Resistance test is performed in four-wire technology. This automatically compensates for the resistance in the test leads.
Using SCHLEICH Kelvin clamps also compensates for the contact resistance at the connection points to the winding. This is the only way to measure resistances below 10 Ω with high precision measurement.
" high voltage test s with AC" (optional)
In addition, you can also use the AC-high voltage test . This is also linked to the device under test via automatic test method switch-over .
» armature test (optional)
armature test is optimally supported by a software wizard. It guides you through the inspection process, ensuring that no bar is overlooked. Manual inspection is performed using test probes or an anchor adapter. Automatic inspection uses a fully automated contact adapter. In this process, either the armature or the contact adapter rotates to make precise and reliable contact with the lamellae. This requires automatic position adjustment built-in. Please contact us for more information. We provide the complete solution tailored to your anchors.
» Rotor test (optional)
The rotor test allows you to locate faults in the stator , or rotor and to detect bar breaks in the squirrel-cage rotor.
motor repair | Manual Check


The MTC2 is ready for immediate use at trade shows. Its unique manual mode allows you to start all test methods (resistance, surge voltage, insulation resistance, and partial discharge testing) directly. There is no need to configure parameters before the tests—simply set your desired TEST VOLTAGE and start the test.
Similar to a multimeter, the software always provides you with the latest results of the test. An evaluation of the winding and the insulation system is thus immediately possible and you can decide whether and what is on the Motor needs to be repaired.
In addition, a large number of engine data can be stored, which can be printed directly on the SCHLEICH standard protocol in conjunction with the measurement results.
motor repair | Manual verification at Squirrel-cage rotor or stator

Special test probes are used to search for the " turn-to-turn fault " in the " stator ." This method can also be applied to the " Squirrel-cage rotor ."
See also: Rotor-stator-Check in SCHLEICHPedia
motor repair | Manual inspection of the rotor

Using special test probes and a armature booster , the rotor is tested using the lamella method. Testing is typically performed between two adjacent lamellae. Quarter-circle or half-circle measurements are also possible. The integrated Ankerassistent software reliably guides the operator through the test.

Here, the armature booster is used for direct measurement on the train.
See also: armature booster in SCHLEICHPedia
motor repair | Fully Automated Testing

In repair applications, a large number of motors and generators can also be tested fully automatically.
The MTC2 offers a fully automatic mode for this purpose, which is used frequently, especially in the " electric motor service and repair ." It can be used in three different ways.
» Predefined Test Plan Routines
The MTC2 comes factory-equipped with a wide variety of test plans for all rated voltage classes. These are predefined to suit a wide range of test tasks and cannot be modified. They can be started immediately in fully automatic mode. The operator simply needs to select the test plan based on the motor’s rated voltage and connect the Motor or stator . Testing can then proceed fully automatically with automatic pass/fail detection.
No technically trained personnel are required to evaluate the measured values!
» Customizable test plans
In fully automatic mode, the test is performed according to a test sequence that can be freelydefined by the operator . The tester evaluates the measurement results fully automatically and displays theresult using a red/green indicator on the monitor. Here, too, no technically trained personnel are required to evaluate the measurement values.
You can create a virtually unlimited number of test sequences that can be adapted and optimized custom to your specific testing situation.
» Trend Analysis – Test Results Comparison
With the integrated comparison of inspections over longer periods of time, the MTC2 provides valuable information to support predictive maintenance. By comparing test results, predictions can be made about the assumed further durability of components and maintenance measures can be initiated at an early stage. The MTC2 helps you reduce downtime and increases the efficiency of your production.
Engine production

The MTC2 can be easily integrated into a production line. The dimensions of the housing are based on a 19" housing, which allows the tester to be seamlessly integrated. The additional interfaces enable complete remote control of the device by a control computer.
For different types of test objects, almost any number of test sequences can be stored in the MTC2. These can be called up and started via an interface.
All test results can be accessed via an interface and can thus be stored in a central memory via the control computer. Additionally, it is possible to save the results locally on the test device or directly on the network at save.
More photos
![]() |
MTC2 on its way to the scene
|
![]() |
Engine testing with the MTC2 portable unit
|
![]() |
Testing a faulty winding
|
![]() |
Investigation of DC Rotors
|
![]() |
Testing Without Limits
|
German PDFs
International PDFs

Do you need personalized product advice?
We'd be happy to provide provide you with comprehensive support and advice on all our products and services.
Give us a call or send us an email!