PortaTest

Insulating oil tester with 60, 80, or 100 kV

Our oil tester PortaTest allows a fully-automatic detection of the dielectric strength of insulating oils.
It is well suited for both applications on-site or in laboratories. Owing to the protection cover that can be opened wide, the test vessel can comfortably be placed on the high-voltage electrodes in the test chamber.

Key-Facts

Indoor. Outdoor. Service.

  • Fully-automatic insulating oil tester
  • Graphical display
  • For use in laboratories
  • For mobile use on site
  • Pre-programmed test sequences according to national and international standards
  • Storage option for up to 10 self-defined test sequences
  • Short switch-off time and power-off < 1 ms avoids carbonisation
  • Average values and single breakdown values displayed
  • Easy user guidance vie 4 function keys
  • Input of sample number and temperature
  • Multilingual software
  • PC-data storage via serial RS232-interface
  • Magnetic stirrer
  • Integrated report printer (plain paper)
  • Numerous test electrodes

Varianten

PortaTest 4018302

  • 100 kV

PortaTest 4018301

  • 80 kV

PortaTest 4018300

  • 60 kV

Beschreibung

Normative Power.

Portatesters are easy to use. Based on the test standard, you select the corresponding test sequence.  Then,  the required test probe and the electrode gap for the selected test standard is displayed. The setting gauge, included in the scope of delivery, facilitates the adjustment of the electrode gap. Afterwards you fill the insulating oil in the test vessel and close the test chamber.

Now the test can be started. The test sequences are run fully-automatic.

The tester contains test sequences for all common national and international standards and test specifications. Furthermore you can self-define and store up to 10 test sequences in the tester. In summary, you can select from following test sequences:

  • 13 standard test sequences cosistent with the following test standards:
    • IEC 60156
    • VDE 0370/96
    • BS 5874/80
    • UTE C27-221/74
    • UNE 21309/89
    • NEN 10156
    • SEV 3141/69
    • CEI 10-1/73
    • ASTM-D 1816/90
    • ASTM-D 877/90
    • JIS 2101/82
    • JIS Si 2101/82
    • IS 6855/92
  • 10 freely-programmable test sequences
  • Short test for fast condition assessment (test duration 5 min)
  • overload test, single breakdown measurement
  • Withstand voltage test (1min … 24h) with programmable rise speed
  • Level test (1-10 levels)

Each oil sample is either marked by the PortaTest with a consecutive number or by the operator by manual number input.

A detailed test report is printed by the installed thermal printer after each test.  Beside the applied test standard the report also contains date, time, sample number, single breakdown values, average value and standard deviation.

If requested, the test results may also be tranferred from the tester to the PC by means of the software PrintCom.  They can also be stored on the PC and test reports may be printed. PrintCom saves the result in the EXCEL®-Format. The EXCEL®-form may be freely- configured by the usesr. Thanks to this, customized test reports may be easily generated.  Further details may be found under our PrintCom software.

Thanks to the clearly arranged graphical display, the operation is very easy.  Four function buttons below the display are sufficiant for the entire operation. They serve to select the test sequence, to enter data, to configure own test sequences and for the tester´s common configuration in the setup, where e.g. the language may be selected.


Technik

Benefits at a glance.

Voltage supply

  • Voltage range 100-250 V, 47-63 Hz (worldwide)
  • key switch for unlocking
  • Fuse protection 6 A

General information

  • Languages: German – English- Italian, may directly be changed in the tester
  • Free programmability of all parameters
  • Fully-automatic measurements
  • Magnetic stirrer
  • Graphical display
  • Integrated clock
  • Integrated printer
  • Integrated interface to communicate with the PC
  • Solid, portable aluminium-housing
  • W 580 mm x D 400 mm x H 440 mm
  • Operating- / and storing temperature 10°…50° C / -20°…+60° C
  • Delivery incl. calibration certificate

Test start

  • Front button to start the test

Safety

  • Test cover safety contact for double circuit safety limit switch – CAT IV

High-voltage test

  • 0-60 kV, 0-80 kV, 0-100 kV
  • R.M.S. measurement
  • voltage rise 0,1-10 KV/s

 

Normative electrodes.

Adaption to standards

  • numerous electrode shapes
  • buffer strips for gap adjustment
  • SCHLEICH GmbH
  • An der Schleuse 11
  • 58675 Hemer | Germany
  • +49 (0) 23 72-94 98-0
  • +49 (0) 23 72-94 98-99
  • info(at)schleich.com